** Flushmate systems not for gravity fed toilets

Sloan Crown Flushometer Troubleshooting Guide


  • 1. Valve does not function

    A. Control stop or main valve is closed.
    Open control stop or main supply valve.
    B. Handle assembly is worn.
    Install Sloan handle repair kit, or replace handle.
    Sloan B-73-A Triple-Seal Flushometer Handle Assembly
    C. Relief valve is worn an dsticking in UP position.
    Replace piston.
    Sloan Crown Piston Kits
  • 2. Insufficient volume of water to adequately siphon fixture.

    A. Control stop not open enough.
    Adjust control stop for desired delivery of water.
    B. Urinal piston parts inside a closet valve.
    Replace piston parts with proper closet piston.
    C. Low consumption valve installed on a non-low consumption fixture.
    Replace with proper flushometer. Crown flushometers are not available with flush volume higher than 1.6 gpf/6.0 Lpf.
    D. Inadequate volume or pressure at supply.
    Increase water pressure or supply (flow) to valve. Consult factory for assistance.
  • 3. Length of flush too short (short flushing) or valve closes off immediately.

    A. Piston assembly is not hand-tight.
    Screw the assembly hand-tight.
    B. Enlarged bypass orifice from corrosion or damage.
    Install NEW inside parts kit to correct problem and update flushometer.
    C. Urinal piston in closet flushometer.
    Replace piston with proper closet piston
    Sloan Crown Piston Kits
    D. Low consumption piston installed in non-low consumption fixture.
    Replace with proper flushometer. Crown flushometers are not available with flush volume higher then 1.6 gpf/6.0 Lpf.
    E. Handle assembly worn.
    Install Sloan handle repair kit.
    Sloan B-73-A Triple-Seal Flushometer Handle Assembly
  • 4. Length of flush to long (long flushing) or fails to close off.

    A. Piston is not seating properly or bypass orifice is clogged because of foreign material, or bypass orifice is clogged by an invisible gelatinous film from "over-treated" water.
    Disassemble the working parts and wash thoroughly. NOTE: Size of the orifice in the bypass is of utmost importance fro the proper metering of water into the upper chamber of the valve. DO NOT enlarge or damage this orifice. Replace Piston if cleaning does not correct problem.
    B. Line pressure has dropped and is not sufficient to force Relief Valve to seat.
    Shut off all control stops until pressure has been restored, then open them again.
    C. CR-125-A main seat is fouled with debris or is worn.
    Clean or replace CR-125-A main seat.
  • 5. Leaking at Handle Assembly

    A. The B-85 seal may have deteriorated.
    Install new B-85 seal. NOTE: The B-85 seal will easily slide inot the bushing if it is wet.
    B. Handle gasket has been omitted
    Install A-31 handle gasket or Sloan handle repair kit.
    Sloan A-31 Handle Gasket
    Sloan V-600-AA Flushometer Vacuum Breaker Tube 1-1/2" x 9"
    C. Valve handle bushing is worn.
    Install Sloan handle repair kit.

Back to top